DC high voltage generator simple "fault" how to rule out?

DC high voltage generator is suitable for DC high voltage test on zinc oxide surge arresters, magnetic discharge lightning arresters, power cables, generators, transformers, switches and other equipment in the power sector, power departments of factories and mines, scientific research units, railways, chemical industry, and power plants. The best new generation of products in the new century.

DC high voltage generator simple "fault" elimination:

1. When the power switch is turned on, the digital display on the control box is off. Please check the power connection and all fuses. If the fuse is damaged, replace the new one with the amperage marked on the side of the fuse holder.

2. If the power switch is turned on, the digital display is on, the high voltage indicator light (green) is off, and no high voltage is added.

1 Fault display "Zero" light does not return to zero position, please turn the high voltage output adjustment knob and fine adjustment knob counterclockwise back to zero to "return to zero" light off, you can add high voltage.

2 Fault display The "other" lights have two phenomena: 1. Grounding protection, 2. Breakage protection a. The grounding wire may not be connected properly, or the ground wire may not be grounded reliably. Please connect a reliable ground, turn off the power, and restart.

b. It may be that the cables connecting the control box and the high voltage tower are not connected properly. After connecting the cables, turn off the power and restart. This fault may occur when the welding heads at both ends of the cable are soldered off. Please solder the soldering tips on both ends of the cable, turn off the power and restart.

C. It is possible to control the disconnection of a wire in the box. Please re-weld, turn off the power and restart.

3, DC high voltage generator is all normal, but in the process of not rising to the rated output voltage protection 1 fault display "overcurrent" light a. sample short circuit, or leakage current is too large.

b. The boost pressure is too fast when large capacitive loads occur. Turn off the power, restart, slow down the boost speed.

2 Fault display "Overvoltage" light is overvoltage protection setting knob is not at the maximum place, please turn clockwise to the maximum place, turn off the power and restart.

4. When short-circuit occurs under high voltage of the test sample, due to the large short-circuit current, it is possible to turn all the fault indicator lights on. This is normal, please restart after troubleshooting the test sample.

5, after troubleshooting the test sample, the fault indicator light is still bright, plus no high pressure.

The power switch is also a fault protection reset switch. Please turn off the power switch and restart it to boost the voltage.

6. If it is found that the voltmeter figures fluctuate greatly, please check whether the 7-pin shield net head and the center shield wire end of the two aviation plugs of the connection cable are disconnected. If it is disconnected, please weld it well.