Mine Stray Current Tester Price Product Quality Commitment

Technical Parameters

Measuring range

measurement accuracy

Note

Current file

0-50mA-250mA
-1A-5A-10A

Exchange ±5%
DC ±2.5%

The pointer must be set to zero before use, that is, zero

Voltage file

0-0.2V-1V
-4V-20V-40V

Mine stray current tester price is a kind of portable rectifier instrument with high sensitivity and multi-range protection circuit. Its characteristics are: a wide range of measurement, a total of 20 ranges, can measure AC and DC stray voltage and stray current, itself does not require power, safe and reliable. It is suitable for testing currents and voltages generated by underground mine rails, water pipes, cables, etc. to prevent electric premature explosion and other explosion accidents caused by stray current discharge, so as to minimize the fire source in the mine.
How to use and precautions
Before measuring, tighten the test bar wiring to the two terminals and adjust the pointer to the zero mark.
The test bars that are connected to the binding posts are placed at the two points to be measured (such as: rails, water pipes, cable jackets, coal, earth, etc.), and the two measuring points should be on different conductors. The measuring point distance is about 2 meters. , Normally measured once every 50-100M, or temporarily determine the test point as needed. When the switch is set to the proper position, the meter indication value is converted to the corresponding current value.
During the measurement, the carbide tip of the measuring probe is in good contact with the measured object, and the meter should be placed horizontally to ensure accurate reading.
When stray currents cannot be measured at one location, DC stray currents should also be tested, and vice versa.
When measuring the stray voltage value, the gear positions are: 50mA corresponds to 0.2V, 250mA corresponds to 1V, 1A corresponds to 4V, 5A corresponds to 20V, and 10A corresponds to 40V. (In other words, the current value multiplied by 4 is the voltage value)
When the amount of stray current is unknown, it should be placed in the maximum range, then gradually reduced until the appropriate measurement file.
When the measured current is greater than 5A, the instrument should only perform short-term measurements. When the measured current is greater than 10A, the instrument should only perform instantaneous measurements. In order to avoid high current burned components and meters.
When measuring DC stray current, such as the head pointer reverse deflection, change the table pen to
The instrument should be placed in the "off" position when not in use.
Product quality commitment
1, the product manufacturing and testing have quality records and test data.
2. For the testing of product performance, we ensure that the product is shipped in a box after inspection according to national regulations and industry standards.

FAN LIGHT

Fan Light,Ceiling Fan with Light,Ceiling Fans with LED Lights,LED Fan Light

Yuyao Flylit Appliance Co.,Ltd , https://www.nbyyflylit.com